The effects upon electromagnetic flowmeter sensitivity of non-uniform fields and velocity profiles

  • Published:
Medical and biological engineering Aims and scope Submit manuscript

Abstract

The magnetic flux density of implantable electromagnetic blood flowmeter can not be made uniform, since the size of flowmeter probe must be very small for physiological reasons. The velocity distribution of blood flow is axially non-symmetrical everywhere in arteries.

Mainly for the above reasons the blood flow can not be measured accurately by implantable electromagnetic flowmeter.

In this paper the relation between induced electromotive force and flow rate for implantable electromagnetic flowmeters is presented, and the errors introduced by the factors mentioned above are discussed. In order to measure pulsatile or axially non-symmetrical flow accurately, appropriate flowmeter probes must be used. Some of these flowmeter probes are shown.

Sommaire

La densité de flux magnétique d'un débitmètre électromagnétique implantable ne peut pas être rendue uniforme, puisque la taille des sondes de débitmètres doit être très petite, pour des raisons physiologiques. La distribution des vitesses du débit sanguin est axialement non-symétrique, en n'importe quel point des artères.

C'est principalement pour les raisons citées ci-dessus que le débit sanguin ne peut pas être mesuré avec précision par les débitmètres électromagnétiques implantables.

L'auteur de cet article présente la relation entre la force électromotrice induite et le débit dans les débitmètres électromagnétiques implantables et y discute les erreurs introduites par les facteurs mentionnés ci-dessus. Pour mesurer le débit pulsé, ou le débit non-symétrique par rapport à l'axe de l'artère, on doit utiliser une sonde de débitmètre appropriée. L'auteur décrit quelques-unes de ces sondes.

Zusammenfassung

Die magnetische Flußdichte eines implantierbaren elektromagnetischen Blutströmungsmeßgerätes kann nicht uniform gehalten werden, da die Größe des Meßkopfes aus physiologischen Gründen sehr klein sein muß. Die Geschwindigkeitsverteilung des Blutstromes ist überall in den Arterien axial unsymmetrisch.

Hauptsäuchlich aus den obengenannten Gründen kann der Blutstrom nicht exakt durch implantierbare elektromagnetische Strömungsmeßgeräte bestimmt werden.

In dieser Arbeit wird die Beziehung zwischen der induzierten elektromotorischen Kraft und der Flußgeschwindigkeit für implantierbare elektromagnetische Strömungsmeßgeräte entwickelt. Die durch die obengenannten Faktoren eingebrachten Fehler werden diskutiert. Um pulsierende oder axial unsymmetrische Strömung exakt zu messen, muß ein geeigneter Meßkopf verwendet werden. Einige solcher Meßköpfe werden demonstriert.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Adler, M (1934) Stroemung in gekruemmtem Rohren.Z. angew. Math. Mech. 14, 257.

    Google Scholar 

  • Cushing, V. (1958) Induction flowmeter.Rev. scient. Instrum. 29, 692.

    Article  Google Scholar 

  • Hale, J. F. (1955) Velocity profiles of oscillating arterial flow.J. Physiol. 128, 629.

    Google Scholar 

  • Kanai, H. (1965) Implantable electromagnetic blood-flowmeter.J. Japan. Soc. Instrum. Control Engrs 4, 328.

    Google Scholar 

  • Kanai, H. (1966) The pulsatile flow measurement by electromagnetic flowmeter: The relation between the accuracy and the uniformity of the magnetic field.Trans. Japan. Soc. Inst. Control Engrs 2, 152.

    Google Scholar 

  • Kanai, H. (1966) An electromagnetic flowmeter for measuring the axially non-symmetrical flow.Trans. Japan. Soc. Inst. Control Engrs 2, 295.

    Google Scholar 

  • Kolin, A. (1945) An alternating field induction flowmeter of high sensitivity.Rev. scient. Instrum. 16, 109.

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Kanai, H. The effects upon electromagnetic flowmeter sensitivity of non-uniform fields and velocity profiles. Med. & biol. Engng. 7, 661–676 (1969). https://doi.org/10.1007/BF02551737

Download citation

  • Received:

  • Revised:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02551737

Keywords

Navigation