Abstract
Objective
To investigate whether a new flow-triggered (FT) system can reduce the patient's inspiratory effort compared to a traditional pressure-triggered (PT) system during weaning from mechanical ventilation.
Design
Prospective study.
Setting
Intensive care unit of a General Hospital.
Patients and participants
10 mechanically ventilated patients, without chronic airway disease, ready to wean.
Measurements
Minute ventilation, breathing pattern, lung mechanics, inspiratory work of breathing (WI) and pressure time product (PTP) of Ppl were obtained in two conditions: 1) unsupported spontaneous breathing through the ventilator circuit (SB); 2) spontaneous breathing with continuous positive airway pressure set at 5 cmH2O (CPAP). Two triggering systems, namely PT and FT, were used in each condition.
Results
Though there was no change in breathing pattern, minute ventilation, and lung mechanics, the magnitude of the inspiratory effort decreased significantly with FT compared to PT in both instances. The added resistance (total flow resistance minus pulmonary resistance) decreased by 37% on average when FT replaced PT. PTP decreased, on average, 27% and 15% during SB and CPAP, respectively, with FT compared to PT (p<0.05). A similar significant decrease was observed in WI.
Conclusion
The new FT system, i.e. flow-by system, reduces the unintentional ventilatory workload upon the patients' inspiratory muscles compared to traditional PT system during weaning from mechanical ventilation.
Article PDF
References
Sassoon CSH, Light RW, Lodia R, Sieck GC, Mahutte CK (1991) Pressuretime product on T-piece, continuous positive airway pressure and pressure support ventilation during weaning from mechanical ventilation. Am Rev Respir Dis 143:469–475
Sassoon CSH, Lodia R, Rheeman CH, Kuei JH, Light RW, Mahutte CK (1992) Inspiratory muscle work of breathing during flow-by, demand flow, and continuous-flow systems in patients with chronic obstructive pulmonary disease. Am Rev Respir Dis 145:1219–1222
Baydur A, Behrakis PK, Zin WA, Jaeger M, Milic-Emili J (1982) A simple method for assessing the validity of the esophageal balloon technique. Am Rev Respir Dis 126:788–791
Polese G, Rossi A, Brandi G et al (1991) Partitioning of respiratory mechanics in mechanically ventilated patients. J Appl Physiol 71:2425–2433
Sassoon CSH, Giron AE, Ely EA, Light RW (1989) Inspiratory work of breathing on flow-by and demand-flow continuous positive airway pressure. Crit Care Med 17:1108–1114
Mead J, Whittemberger JL (1953) Physical properties of human lungs measured during spontaneous respiration. J Appl Physiol 5:770–796
Sassoon CSH (1992) Mechanical ventilator design and function: the trigger variable. Respir Care 37:1056–1069
Cox D, Tinloi SF Farrimond JG (1988) Investigation of the spontaneous modes of breathing of different ventilators. Intensive Care Med 14:532–537
Field S, Sanci S, Grassino A (1984) Respiratory muscle oxygen consumption estimated by the diaphragm pressuretime index. J Appl Physiol 57:44–51
Petrof BJ, Legare M, Goldberg P, Milic-Emili J, Gottfried SB (1990) Continuous positive airway pressure reduces work of breathing and dyspnea during weaning from mechanical ventilation in severe chronic obstructive pulmonary disease. Am Rev Respir Dis 141: 281–289
Katz JA, Marks JD (1985) Inspiratory work with and without continuous positive airway pressure in patients with acute respiratory failure. Anesthesiology 63:598–607
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Polese, G., Massara, A., Poggi, R. et al. Flow-triggering reduces inspiratory effort during weaning from mechanical ventilation. Intensive Care Med 21, 682–686 (1995). https://doi.org/10.1007/BF01711550
Received:
Accepted:
Issue Date: