Summary
A technique is presented for recording both local blood flow, by means of hydrogen clearance method, and neuronal activity of the cat's cortex using a single microelectrode. The requirements of electrodes for recording cerebral blood flow are different from those for recording extracellular spikes. Only electrodes within a small range of impedance (measured at 10 Hz) are suitable to record both. A bridge method is used for the recordings. The bridge signal is split in order to amplify cell activities and H2 responses separately. Three examples demonstrate the method and the importance of appropriately selected electrodes.
Similar content being viewed by others
References
Aukland, K., Bower, B. F., Berlinger, R. W.: Measurement of local blood flow with hydrogen gas. Circulation Res.14, 164–187 (1964)
Gesteland, R. C., Howland, B., Lettvin, J. Y., Pitts, W. H.: Comments on microelectrodes. Proc. IRE 1856–1862 (1959)
Green, J. D.: A simple microelectrode for recording from the central nervous system. Nature182, 962 (1958)
Guld, Ch.: A glass covered platinum microelectrode. Med. Electron. Biol. Engng.2, 317–327 (1964)
Heiss, W.-D., Hayakawa, T., Waltz, A. G.: Cortical neuronal function during ischemia. Effects of occlusion of one middle cerebral artery on single unit activity in cats. Arch. Neurol. (Chicago) in press (1976)
Hyman, E. S.: Linear System for quantitating hydrogen at a platinum electrode. Circulation Res.9, 1093–1097 (1961)
Pasztor, E., Symon, L., Dorsch, N. W. C., Branston, N. M.: The hydrogen clearance method in assessment of blood flow in cortex, white matter and deep nuclei of baboons. Stroke4, 556–567 (1973)
Rappelsberger, P.: Registrierung von Aktionspotentialen im Gehirn. Diplomarbeit, Technische Universität Wien (1968)
Robinson, D. A.: The electrode properties of metal microelectrodes. Proc. IEEE56, 1065–1071 (1968)
Wolbarsht, M. L., MacNichol, E. F., Wagner, H. G.: Glass insulated platinum micro-electrode. Science132, 1309–1310 (1960)
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Rappelsberger, P., Heiss, W.D., Vollmer, R. et al. A technique for recording local blood flow and neuronal activity with a single microelectrode. Pflügers Arch. 369, 183–186 (1977). https://doi.org/10.1007/BF00591575
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF00591575